up

HOME SERVICE SITE MAP CONTACT US
This autoclave is made by incorporating a high performance vacuum pump into the HV series autoclave and equipped with fully automatic drying function.
Autoclave with full automatic drying function
HICLAVE
HVP series
HVP-50
  • Clean drying without contamination
  • Pre-vacuum air removal using a vacuum pump
  • Electro-mechanical lid lock system
  • Dual interlock mechanism

  •  ■ Features & Functions
    Clean drying without contamination
    After sterilization, drying is performed with the lid closed.
    Clean drying can be carried out without worry of contamination.
    Pre-vacuum air removal using a vacuum pump
    A vacuum pump is installed to remove residual air with vacuum.
    This is effective when sterilizing objects put in containers or bags from which air is difficult to escape.
    Electro-mechanical lid lock system
    The lid can be opened and closed easily and reliably with the one-touch lever.
    A pressure seal method is used as a safety measure.
    Dual interlock mechanism
    The lid is locked by dually detecting the chamber pressure and chamber temperature during operation.
    The lock state is maintained even if an operation is manually interrupted.
    Memory function
    Set contents can be registered for each mode.
     ■ Options for upgrading a system
      
    Recorder
    (Image is a chart type recorder.)
    ・Chart type recorder: 6-point dotting
    ・Hybrid recorder: 8-point dotting
    ※Chart width: 100 mm
    ※Class 2 T-type thermocouple included

    Required utilities   
    Power supply:Refer to the specifications. Household outlets are not recommended.
    ※Ask us for details.
    Water supply facility:For water seal vacuum pump (Supply water using a hose.)
    Drainage:Steel pipes are recommended for high temperature high pressure drainage.

    ※Ask us for details.
     ■ Display panel
     ■ Operation cycle of each mode
      MODE 1 : FABRICVacuum air removal
    (Vacuum drawingx2)
    HeatingSterilizationExhaust,
    Drainage
    Drying
      ●Main applications: Sterilization of gauze, cloths that require drying after sterilization.
      MODE 2 : FABRICVacuum air removal
    (Vacuum drawingx3)
    HeatingSterilizationExhaust,
    Drainage
    Drying
      ●Main applications: Sterilization of gauze, cloths that require drying after sterilization.
      MODE 3 : SOLIDVacuum air removal
    (Vacuum drawingx2)
    HeatingSterilizationExhaust,
    Drainage
      ●Main applications: Sterilization of instruments made of glass, ceramic, metal or rubber that can withstand high temperature, high pressure steam and rapid pressure reduction during exhaust.
      MODE4 : LIQUIDHeatingSterilizationExhaust
      ●Main applications: Sterilization of liquid, which can withstand high temperature and high pressure steam such as water, culture medium, reagent, reagent solution, liquid medicine, etc., and sterilization of waste using sterilization bag.
      MODE5 : ADDITIONAL DRYINGDrying
      ●Main applications: Additional drying of gauze, cloths
     ■ Option
                    
    Stainless steel round wire basket Sterilization bucket, stainless steelRound dressing drumstainless steel
    φ270 x H220 mm
    φ270 x H270 mm
    φ270 x H220 mm
    φ270 x H270 mm
    [Small]
    φ170 x H120 mm

    [Medium]
    φ200 x H150 mm

    [Large]
    φ260 x H180 mm
                                                   
    Sterilization tape, Sterilization label, Sterilization bag
    ■Sterilization tape: Clear color change makes it easy to identify exposure to sterilization process.
    ■Sterilization label: Paste directly on the object to be sterilized, you can check the sterilization treatment. After sterilization, the letters "sterilized" appear and it can be easily distinguished from untreated materials.
    ■Sterilization bag: There are sterilization bags that are made of paper, roll type or one seal back type. Most of them are heat seal type, but tape seal type and aluminum plate seal type can be also available.

    ※Ask us for prices and more details.
     ■ Specifications
    Product nameHICLAVE with a vacuum pump
    ModelHVP-50
    Effective capacity50 L
    Chamber size (φ x H mm)300 x 710
    Operation mode5 modes
    (FABRICS with vacuum drawing 2 times, FABRICS with vacuum drawing 3 times, SOLID, LIQUID, ADDITIONAL DRYING)
    Sterilization temperature setting range105℃ - 135℃ Variable
    Lid lock97℃, fixed
    Time setting
    range
    Sterilization1min - 250 min, Remaining time is displayed
    Drying1 min - 90 min, Remaining time is displayed
    Exhaust setting3 levels (Without exhaust, Fine exahut, Small amount of exhaust)
    Drying methodVacuum drying using a water seal vacuum pump and heating of the outside of the chamber
    Maximum working pressure0.255MPa
    ThermometerDigital display 5℃ - 137℃
    Pressure gaugeAnalog display -0.1MPa - 0.4MPa
    Safety devicesPressure safety valve, Earth leakage breaker, Low water cut off device, Earth leakage breaker for vacuum pump
    Alarms, error displayLow water heating, Temperature sensor wire disconnection, Over-temperature, Over-cool, Over-pressure,
    Abnormality in lid opening / closing lever lock, Lack of water for vacuum pump bottle
    Outside dimensions (W x D x H mm)540 x 630 x 1040
    Weight (Approximately)80 kg
    Pressure vessel categorySmall sized pressure vessel
    Chamber materialStainless steel (SUS304)
    Power supplyAC120V / 220V / 230V / 240V, 50Hz / 60Hz(Single phase)
    (Specify voltage when ordering.)
    Required power supply facility 2.0kW
    Power connection 
    Ring terminal
    Supplied accessoriesBottom plate, Flexible drainage hose, Drainage hose, Water supply hose, Caster stopper (2)
    ※Specifications and appearance are subject to change without notice due to continuous product improvement.