A multi-use model equipped with a sterilization mode
for fragile medicine bottles or retort packs.
Compact sterilizer for
medicine bottles or retort packs

HLM series
HLM-36ELB

  • Pressure control method
  • A cooling device that can prevent damage to containers.
  • Electro-mechanical lid lock system Ⅱ
  • Dual interlock mechanism

  •  ■ Features & Functions
    Pressure control method
    High temperature and high pressure sterilization is performed by gradually raising the chamber temperature and pressure. This is effective for preventing damage to aluminum packs or film packs for retort foods.
    A cooling device that can prevent damage to containers.
    Cooling water is sprayed while pressurizing the chamber with air in order to prevent rapid pressure change due to sudden condensation of vapor in the chamber during shower cooling. The seal portion of a bottle and container is prevented from being damaged.
    Automatic operation
    All stages from water supply, pressurization, sterilization (disinfection) to cooling are performed automatically.                
    Compact body made of stainless steel
    A stainless steel body that is clean and excellent in corrosion resistance is used.
    A compact size autoclave which does not take space.
    Electro-mechanical lid lock system Ⅱ
    The electric clamp lid is employed. Opening/locking/unlocking of the lid can be performed simply by pressing a button.
    The chamber opening can be widely opened for easy loading and unloading.
    Dual interlock mechanism
    The lid is locked by dually detecting the chamber pressure and chamber temperature during operation.
    The lock state is maintained even if an operation is manually interrupted.
    Memory function
    Set contents can be registered for each mode.
    Programmable auto-start function
    An operation starts at the set time.
    Sufficient treatment capacity (Loading capacity)

    Pouch volumePouch loading capacity
    250~300 cc 30~45 P
    400~500 cc 15~30 P
    1000 cc 15 P
    (reference)
     
    Equipped with external output terminal as standard
    As a printer output connector is equipped as standard, it is easy to output data.
    Installation requirements are minimal and maintenance cost is low
    It can be installed simply by providing power source and water supply/drainage. See the following installation requirements.Qualification for handling boilers or pressure vessels is unnecessary. (It is necessary for users to inspect it once a year.)
    Required utilities   
    Power supply : Ask us for details.
    Water supply facility:Water supply pressure of 0.23 to 0.40MPa, Flow rate of6ℓ /min or more
    Drainage:Steel pipes are recommended for high temperature high pressure drainage.

    ※Ask us for details
     ■ Display panel
     ■ Operation cycle of each mode
      MODE1:  Sterilization
          (Natural cooling)
    Automatic
    water supply
    HeatingSterilizationNatural coolingWarming
      ●Main applications: Sterilization of glass, ceramic and metal instruments that can withstand high temperature and high pressure steam
      MODE 2:  Sterilization
           (Shower cooling)
    Automatic
    water supply
    HeatingSterilizationShower coolingDrainage
      ●Main applications: Sterilization of glass, ceramic and metal instruments that can withstand high temperature and high pressure steam
                        Sterilization of aluminum packs or film packs ( Without pressure control)
       
      MODE 3:  Sterilization
           (Shower cooling)
    Automatic
    water supply
    Heating PressurizationSterilization (Pressure cooking)Shower coolingDrainage
      ●Main applications: Sterilization of aluminum packs or film packs ( With pressure control)
     ■ Options for upgrading a system (retrofittable)
                                                                                                                  
    Floating sensorPrinter
    A sterilization timer operates after the temperature of the materials to be sterilzed reaches the set temperature.Data collection (Thermal recording type)
     ■ Option
                    
    Stainless steel round wire basket Sterilization bucket, stainless steelTray, stainless steel
    φ330 x H195(mm)
    φ330 x H270(mm)
    φ330 x H270(mm)

    [ Tray 38]
    φ320 x H38(mm)

    [ Tray 48]
    φ320 x H48(mm)
                                                   
    Sterilization tape, Sterilization label, Sterilization bag
    ■Sterilization tape: Clear color change makes it easy to identify exposure to sterilization process.
    ■Sterilization label: Paste directly on the object to be sterilized, you can check the sterilization treatment. After sterilization, the letters "sterilized" appear and it can be easily distinguished from untreated materials.
    ■Sterilization bag: There are sterilization bags that are made of paper, roll type or one seal back type. Most of them are heat seal type, but tape seal type and aluminum plate seal type can also be available.

    ※Ask us for prices and more details.
     ■ Specifications
    Product nameRetort Sterilizer HLM
    ModelHLM-36ELB
    Effective capacity63L
    Chamber size (φ x H mm)360 x 620
    Operation mode3 modes (MODE 1:Sterilization and natural cooling, MODE 2:Sterilization and shower cooling,MODE 3:Sterilization and shower cooling)
    Sterilization temperature setting range105-134℃ (MODE 1), 105-125℃ (MODE 2 and 3) Variable
    Lid lock97℃, fixed
    Time setting
    range
    Sterilization1-250 min, Remaining time is displayed
    cooling10~250 min, Remaining time is displayed
    Maximum working pressure0.235MPa
    ThermometerDigital display5~136℃
    Pressure gaugeAnalog display 0~0.4MPa
    Safety devicesPressure safety valve, Over-current and earth leakage breaker, Low water cut off device
    Alarms, error displayLow water heating, Temperature sensor wire disconnection, Over-temperature, Over-pressure, Lid lock abnomality
    Outside dimensions (W x D x H mm)602 x 679 x 1064
    Weight (Approximately)120 kg
    Pressure vessel categorySmall sized pressure vessel
    Chamber materialStainless steel (SUS304)
    Power supplyAC200V, 50/60 Hz (Single phase)
    Required power supply facility 5.0kW
    Power connection
    (Cord length)
     
    Ring terminal(2.7m)
    Supplied accessoriesStainless steel wire basket (3), Bottom plate (1)